topnavREV.htmltopnavREV.html
 |

| Type: |
Motor-driven, compact interchangable transducers |
| Transducer Frequency: |
35 MHz and/or 50 MHz |
Scanning
Method: |
Variable field sector scanner |
| Sector angle: |
38º or 20º fields |
Scanning
speed: |
Variable 12.5 frames per second (fps) |
| Observable Range (Variable): |
18.5 mm wide x 14 mm deep in 38º field
12.0 mm wide x 14 mm deep in 20º field
|
| Display: |
Dual screen simultaneous display with live zoom and standard screen display |
| In 38° field sector: |
Axial: 0.027 mm
Lateral: 0.035 mm |
| In 20° field sector: |
Axial: 0.027 mm
Lateral: 0.023 mm |
|
ACOUSTIC AXIAL RESOLUTION |
| 35 MHz Transducer: |
0.068 mm |
| 50 MHz Transducer: |
0.050 mm |
| Gain Curves: |
Logarithmic with user-selectable window (contrast) and level (brightness) control |
| Distance Measurements |
Angle-to-angle, sulcus-to-sulcus, corneal thickness, and scleral thickness |
| Anterior Segement Biometry: |
Single measurement of cornea thickness, anterior chamber depth (ACD), and lens thickness
A-scan profile with two markers / dual caliper measurements
Angle in degrees |
| Scan Time: |
45 seconds (depending on RAM) |
| Recording Frame Rate: |
12.5 or 25 fps |
| Voltage: |
90 - 240 VAC |
| Frequency: |
50/60 Hz |
| Maximum Current: |
6.5 A (low voltage range) or 3.5 A (high voltage range)
|
| Type: |
Liquid crystal display (LCD) flat panel monitor |
| Viewable Size: |
15” |
| Resolution: |
1024 x 768 |
| Pixel Pitch: |
0.297 mm |
| Contrast Ratio: |
450:1 |
| Computer: |
17" H x 8.9" W x 18.4" D |
| Monitor: |
15.8" H x 16.1" W x 6.7" D |
| Computer: |
42 lbs |
| Monitor: |
11.5 lbs |
|
|